| Model |
D501 |
D1001 |
D1201 |
D1601 |
|
|
| Temperature Range |
-50 to 550 ºF (-45 to 287 ºC) |
0 to 1000 ºF
(-18 to 540 ºC) |
186 to 1207 ºF
(86 to 653 ºC) |
186 to 1600 ºF
(86 to 871 ºC) |
|
|
| Relative Humidity Measurement |
Kit Included |
No |
|
|
|
|
| Emissivity Adjustment |
Automatic Emissivity Compensation System |
|
|
| Calibration Requirement |
None |
|
|
| Linearity Error (% of Reading) |
± 1 % |
± 3 % |
± 3 % |
± 3 % |
|
|
| Emissivity Error |
± 1 % maximum of difference between target temperature and instrument temperature when touching, for emissivity of 0.8 to 1.0 |
|
|
| Repeatability |
± 0.1 ºF (± 0.1 ºC) |
|
|
| Resolution |
± 0.1 ºF (± 0.1 ºC) |
|
|
| Display/Update |
Bright LEDs at 10 times per second |
|
|
| Response Time |
80 msec approximately |
|
|
| Field of View |
1:1 (53º approximately) |
|
|
| Minimum Spot Size |
.25" (6.4 mm) approximately |
|
|
| Spectral Sensitivity |
2 to 20 microns |
|
|
| Analog Output* |
1 Mv/ºF (1 mV/ºC) Optional |
|
|
| F/C Conversion |
Yes |
Yes |
Yes |
No |
|
|
| Remote Sensor** |
Available on all models' standard on D1201 and 1601 |
|
|
| Instrument Operating Temperature |
32 to 122 ºF (0 to 50 ºC) |
|
|
| Battery Life |
Approximately 5000 readings from one 9V alkaline battery |
|
|
| Dimensions |
Main Case: 3.375" x 5" x .75" (8.5 x 12.5 x 2 cm) |
|
|
| Weight |
7 oz. |
7 oz. |
N/A |
N/A |
|
|
| Weight - Remote Sensor Version |
9 oz. |
9 oz. |
9 oz. |
9 oz. |