Exergen Corporation
 
Industrial Division
SmartIRt/c
 
SmartIRt/c
Plug and play, linearized, precise accuracy, with integral design which eliminates thermal drift. No leakage current, impedance, or cold junction compensation to worry about.
 
SmartIRt/c Ten Compelling Reasons for Accuracy

SmartIRt/c datasheet (SmartIRt/c.3, SmartIRt/c.5, SmartIRt/c.20, SmartIRt/c.40)

 
SmartIRt/c Installation Instructions
SmartIRt/c Engineering notes
Smart-microIRt/c Data Sheet
 
The SmartIRt/c’s are CE compliant,click here for the full test report.
WARNING - 4-20mA models will only drive 50 ohm impedance, if your input device exceeds 50 ohms, please select a voltage output. Models which will drive over 250 ohms are available via special request.

WIRING INSTRUCTIONS AND ALARM CODES:

Complete guide to wiring the SmartIRt/c with alarm codes for troubleshooting.
Threshold Pin Option - The Threshold Pin provides the user with an independent line that goes high (over 5V) whenever the unit has reach the customer selected state. This feature is most commonly used in applications were immediate shut-off or operator intervention is required when a certain temperature is reached.
Alarm Pin Option - The Alarm Pin provides the user with an independent line that goes high (~ 5V) whenever the unit is in an error state.* This feature is useful in applications were immediate "shut-off" is required. There are also applications were the controller cannot determining when the signal line is showing an error message and an independent line is required.
TA PULL DOWN PIN OPTION - The TA Pull Down Pin allows the user to select whether the Smart IRt/c™ output the target temperature or the ambient temperature. This is achieved by pulling and holding the pin to ground. For as long as the pin is held low, plus one response time (less then 250 msec), the unit will output ambient temperature, releasing the pin will return the unit to normal operation

SMARTIRt/c ASSEMBLY DRAWING: Click here for the latest mechanical assembly drawing

Pricing Information
To obtain pricing information, please contact your local distributor, or call the factory at (617) 923-9900.  U.S. distributors | International distributors
Factory Experts are always at your service. Contact 617.923.9900, ext 6238 or 6263,or email: industrial@exergen.com, an expert will be in touch with you within 1 hour (during factory hours of 8AM - 5PM, US EST).
400 Pleasant St. Watertown, MA 02472
Toll Free : 1-800-422-3006
Phone : 617-923-9900 Fax : 617-923-9911