|
An occasional problem introduced by
switching-type thermocouple dataloggers is signal offset
caused by the switching transient. The IRt/c is a completely
passive device and produces an electrical signal entirely
via thermoelectric effects, but does contain both resistance
and capacitance above the levels found with conventional
thermocouples. Many interface devices generate a small
leakage current, which induces no shift in signal with
conventional low impedance (<100 ohm) thermocouples, but
may induce an offset with the higher IRt/c impedance (~3K
ohm). This type of offset is normally stable and is simply
calibrated out by adjusting the device’s OFFSET or ZERO
adjustment.
However, switching the thermocouple
input can also cause offsets in IRt/c readout due to the
presence of capacitance, if the signal leads are connected
in a differential fashion to the amplifier input. A
switching transient voltage stores a charge in the
capacitance, which can cause the equivalent of leakage
current offset. This offset could also be calibrated out,
but may not be stable. A preferred method is simply to
ground the negative side of the t/c input as shown.
The ground provides a path for the
charge caused by the switching transient to dissipate, thus
eliminating the offset. The twisted shielded pair wire with
shield connected to ground will compensate for any loss of
noise rejection, and thus provide a clean signal.

|