All infrared-based sensing systems must be calibrated for
specific material surface properties (for example, the
amount of heat radiated from the target surface,
environmental heat reflections, etc.). This calibration is
performed by measuring the target surface temperature with a
reliable independent surface temperature probe. The easiest
and fastest method of accurately calibrating out these
effects is to use an Exergen Microscanner D-Series hand-held
Infrared Thermometer with a patented Automatic Emissivity
Compensation System, which gives a true reading regardless
of emissivity. Your authorized IRt/c Distributor will be
pleased to make a D-Series available for your installation.
To calibrate Adjustable models (IRt/c.xxA) see Tech Note No.
60.
The following procedure is recommended:
- Install the IRt/c as close as practical to view the
target material to be measured.
- Wire the IRt/c to the controller,
PLC, transmitter,
etc. in standard fashion (including ground shield as in Tech
Note #82). As with conventional thermocouples, red wire is
always (-).
- Bring the process up to normal operating temperature
and measure the actual temperature of the target material
with the Microscanner D-Series Infrared Thermometer.
- Adjust "input offset," "zero,"
"low cal," on the readout device to match the
Microscanner reading.
Installation Complete. (For OEM installations preset the
same adjustments. Individual calibration is not required.)

|